EDAN iM50 User Manual page 74

Patient monitor
Hide thumbs Also See for iM50:
Table of Contents

Advertisement

Patient Monitor User Manual
Message
IBP Catheter Off
IBP Sensor Err
YY Comm Fail (YY
stands for the IBP
label name: Art, PA,
CVP,
RAP,
LAP,
ICP, P1 and P2)
C.O.
C.O. Comm Fail
C.O. TI Sensor Off
C.O. TB Sensor Off
C.O. TEMP Out Of
Range
AG
AG Comm Fail
CO
Out Of Range
2
N
O Out Of Range
2
Cause
IBP catheter falls off due to
patient movement.
Malfunction in the IBP
sensor or in the extension
cable.
IBP
module
failure
communication failure
C.O.
module
failure
communication failure
C.O.
TI
sensor
connected
C.O.
TB
sensor
connected
TI/TB measuring value is
beyond measuring range.
AG
module
failure
communication failure.
The
CO
concentration
2
exceeds the accuracy range
of AG module.
The
N
O
concentration
2
exceeds the accuracy range
of AG module.
Alarm Level
High
Medium
or
High
or
High
not
Low
not
Low
High
or
High
High
High
- 62 -
Alarm Information
Action Taken
Check the catheter
connection
and
reconnect it.
Replace
the
IBP
sensor
or
the
extension cable.
Stop
measuring
function
of
IBP
module, and notify
biomedical engineer
or
Manufacturer's
service staff.
Stop measuring of
C.O.
module,
or
notify
biomedical
engineer
or
Manufacturer's
service staff.
Insert
injective
temperature sensor.
Insert TB sensor.
Please check TI/TB
sensor.
Stop
measuring
function
of
AG
module, and notify
biomedical engineer
or
Manufacturer's
service staff.