EDAN iM50 User Manual page 78

Patient monitor
Hide thumbs Also See for iM50:
Table of Contents

Advertisement

Patient Monitor User Manual
Message
CO
Check Adapter
2
CO
Sensor
Over
2
Temp
CO
Out Of Range
2
EtCO
Overrange
2
FiCO
Overrange
2
CO
Sensor Faulty
2
CO
Occlude
2
CO
Replace Adapter
2
Cause
1) For the Respironics CO
module: The cannula is off
or disconnected.
2) For the EDAN CO
module: The water trap is
disconnected or not properly
connected.
CO
sensor
temperature
2
exceeds +40 °C.
The
CO
concentration
2
exceeds the accuracy range
of CO
module.
2
The EtCO
concentration
2
exceeds the measurement
range.
The FiCO
concentration
2
exceeds the measurement
range.
CO
module failure
2
Water trap of SideStream is
occluded.
Replacement of the adapter
is required.
Alarm Level
2
Low
2
High
High
High
High
High
High
Medium
- 66 -
Alarm Information
Action Taken
1)
For
the
Respironics
CO
module:
Check
whether the adapter
is properly connected
or
replace
the
adapter.
2) For the EDAN
CO
module:
2
Properly connect the
water trap.
Stop
using
measuring
function
of
CO
module,
2
notify
biomedical
engineer.
Reduce
CO
concentration.
Please
check
the
monitor or patient
status and adjust the
gas
concentration
accordingly.
Stop
measuring
function
of
CO
module,
notify
biomedical engineer.
Make sure the gas
exhaust works well
Replace the adapter.
2
2
2