Mode
Measuring Interval in AUTO
Mode (unit: minute)
Continuous
Measuring Parameter
Pressure Unit
☆ Measuring Range
☆ Adult Mode
☆ Pediatric Mode
☆ Neonatal Mode
☆ Alarm Type
☆ Cuff
Pressure
Range
Pressure Resolution
☆ Maximum Mean Error
☆ Maximum Standard Deviation
Maximum Measuring Period
Adult/Pediatric
Neonate
Typical Measuring Period
Dual Independent Channel Overpressure Protection
Adult
Pediatric
Neonatal
Pre-inflation Pressure
Manual, Auto, Continuous, Sequence
1/2/3/4/5/10/15/30/60/90/120/180/240/360/480 and User
Define
5 min, interval is 5 s
SYS, DIA, MAP, PR
kPa, mmHg, cmH
SYS: 25 mmHg to 290 mmHg
DIA: 10 mmHg to 250 mmHg
MAP: 15 mmHg to 260 mmHg
SYS: 25 mmHg to 240 mmHg
DIA: 10 mmHg to 200 mmHg
MAP: 15 mmHg to 215 mmHg
SYS: 25 mmHg to 140 mmHg
DIA: 10 mmHg to 115 mmHg
MAP: 15 mmHg to 125 mmHg
SYS, DIA, MAP
0 mmHg to 300 mmHg
Measuring
1 mmHg
±5 mmHg
8 mmHg
120 s
90 s
20 s to 35 s (depend on HR/motion disturbance)
(297±3) mmHg
(245±3) mmHg
(147±3) mmHg
O
2
- 251 -