EDAN iM50 User Manual page 271

Patient monitor
Hide thumbs Also See for iM50:
Table of Contents

Advertisement

Patient Monitor User Manual
A.12 CO
2
Complies with ISO 80601-2-55: 2011.
EDAN G2 Sidestram Module
Intended
Adult, pediatric, neonatal
Patient
Measure
EtCO
Parameters
Unit
mmHg, %, kPa
EtCO
☆ Measuring
FiCO
Range
AwRR 2 rpm to 150 rpm
EtCO
Resolution
FiCO
AwRR 1 rpm
EtCO
☆ Accuracy
AwRR ± 1 rpm
Drift of
Measure
Meets the requirements of the measure accuracy
Accuracy
Sample
Gas
70 ml/min or 100 ml/min (optional), accuracy: ± 15 ml/min
Flowrate
Warm-upTime Display reading within 20 s; reach to the designed accuracy within 2 minutes.
< 400 ms (with 2 m gas sampling tube, sample gas flowrate: 100 ml/min)
Rise Time
< 500 ms (with 2 m gas sampling tube, sample gas flowrate: 70 ml/min)
Response
< 4 s (with 2 m gas sampling tube, sample gas flowrate: 100 ml/min)
, FiCO
, AwRR
2
2
0 mmHg to 150 mmHg (0 % to 20%)
2
0 mmHg to 50 mmHg
2
1 mmHg
2
1 mmHg
2
± 2 mmHg, 0 mmHg to 40
mmHg
± 5% of reading, 41 mmHg to
70 mmHg
± 8% of reading, 71 mmHg to
2
100 mmHg
± 10% of reading, 101 mmHg
to 150 mmHg
± 12% of reading or ± 4 mmHg,
whichever is greater
Typical conditions:
Ambient temperature: (25 ± 3) ° C
Barometric pressure: (760 ± 10) mmHg
Balance gas: N
Sample gas flowrate: 100 ml/min
All conditions
- 259 -
Product Specification
2