Do not place the module or accessories in any position that might cause it to fall on
the patient.
16
Only use Masimo authorized devices with the module. Using unauthorized devices
with the module may result in damage to the device and/or patient injury.
17
Carefully route the sampling line to reduce the risk of patient entanglement or
strangulation.
18
Do not lift the module by the NomoLine capnography sampling line as it could
disconnect from the module, causing the device to fall on the patient.
19
Do not use the module during magnetic resonance imaging (MRI) or in an MRI
environment.
20
Only use sample lines intended for anesthetic agents if N
are being used.
21
Do not re-use disposable single-patient use NomoLine Family sampling lines due to
the risk of cross contamination.
22
Do not use the NomoLine Infant/Neonate Airway Adapter Sets for adults/pediatrics as
they may cause excessive flow resistance (0,7 ml dead space).
23
Do not use the NomoLine Adult/Pediatric Airway Adapter Sets for infants/neonates as
the airway adapter adds 6 ml dead space.
24
Do not apply negative pressure to remove condensed water from the NomoLine
Family sampling line.
25
The module is not intended to be used for returning exhaust gases to the patient
circuit. Exhaust gases should be returned to a scavenging system.
26
Disconnect the device from AC mains by removing the device cable connection from
the host device.
27
The module should not be used as the sole basis for medical decisions. It must be
used in conjunction with clinical signs and symptoms.
28
Use of high-frequency electrosurgical equipment in the vicinity of the module may
produce interference and cause incorrect measurements.
29
Do not use the module with metered-dose inhalers or nebulized medications as this
may clog the bacteria filter.
30
Properly apply sampling lines according to the sampling lines directions for use.
Misapplied sampling lines that become partially dislodged may cause no or incorrect
readings.
31
Replace the sampling line if the sampling line input connector starts flashing red, or
host device displays a check sampling line type of message.
WARNING
- 153 -
Monitoring CO
O and/or anesthetic agents
2
2