EDAN iM50 User Manual page 273

Patient monitor
Hide thumbs Also See for iM50:
Table of Contents

Advertisement

Patient Monitor User Manual
Halothane
Enflurane
Isoflurane
Sevoflurane
Xenon
Hehelium
Metered dose inhaler propellants
Desflurane
Ethanol
Isopropanol
Acetone
Methane
Respironics Sidestram and Mainstram Modules
Applicable Patient Type
Technique
Measure Parameters
Unit
☆ Measuring Range
☆ EtCO
2
☆ FiCO
2
☆ AwRR
Resolution
☆ EtCO
Accuracy
2
4%
5%
5%
5%
Not applicable
Not applicable
Not applicable
15%
0.1%
0.1%
0.1%
1%
Adult, pediatric and neonatal patients
Infra-red Absorption Technique
EtCO
, FiCO
, AwRR
2
2
mmHg, %, Kpa
0 mmHg to 150 mmHg
3 mmHg to 50 mmHg
0 rpm to 150 rpm (Mainstream)
2 rpm to 150 rpm (Sidestream)
EtCO
1 mmHg
2
FiCO
1 mmHg
2
AwRR
1 rpm
± 2 mmHg, 0 mmHg to 40 mmHg
± 5 % of reading, 41 mmHg to 70 mmHg
± 8 % of reading, 71 mmHg to 100 mmHg
± 10 % of reading, 101 mmHg to 150 mmHg
± 12% of reading, RR is over 80 rpm (sidestream)
There will be no degradation in performance due to
- 261 -
Product Specification
None
None
None
None
Not applicable
Not applicable
Not applicable
None
None
None
None
None

Advertisement

Table of Contents
loading

This manual is also suitable for:

Im60Im70Im80M50M80

Table of Contents