Enflurane
Isoflurane
Sevoflurane
Xenon
Helium
Desflurane
Barometric Pressure on EtCO
Quantitative Effect
Ambient Barometric, Operational
(0 – 40) mmHg: ± 1 mmHg additional error
(41 – 70) mmHg: ± 2.5% additional error
(71 – 100) mmHg: ± 4% additional error
(101 – 150) mmHg: ± 5% additional error
*Additional worst case error when compensation for P
is correctly selected for the actual fractional gas constituents present.
NOTE:
Respiration Rate accuracy was verified by using a solenoid test setup to deliver a square
wave of known CO
used. Respiration rate was varied over the range of the device. Pass/Fail criteria was
comparison of the respiratory rate output from the sensor to the frequency of the square
wave.
Masimo Sidestream Module
Dimensions
Weight
According to the degree of
protection
against
ingress of water
5
5
5
80
50
15
Measurement Values:
2
concentration to the device. 5% and 10% CO
2
91 mm × 60 mm × 37 mm (3.6 × 2.4 ×1.4), exclduing
cable, tubing and Nomoline.
≤ 98 g, exclduing cable, tubing and Nomoline.
IP34
harmful
(41 – 70) mmHg: ± 2.5% additional error
(71 – 100) mmHg: ± 4% additional error
(101 – 150) mmHg: ± 5% additional error
*Additional worst case error when compensation
for P
, O
, N
O, anesthetic agents, or helium is
B
2
2
correctly selected for the actual fractional gas
constituents present.
Desflurane:
The presence of desflurane in the exhaled breath at
concentrations greater than 5% will positively bias
Carbon Dioxide values by up to an additional 3
mmHg at 38 mmHg.
Xenon:
The presence of Xenon in the exhaled breath will
negatively bias Carbon Dioxide values by up to an
additional 5 mmHg at 38 mmHg.
, O
, N
B
2
2
- 263 -