Patient Monitor User Manual
☆ Adult Mode
☆ Pediatric Mode
☆ Neonatal Mode
☆ Alarm Type
Pressure Resolution
☆ Maximum mean error
☆ Maximum standard deviation
Maximum measuring period
Adult
Pediatric
Neonate
Overpressure protection
Adult/Pediatric
Neonate
Pre-inflation Pressure
Adult Mode
Pediatric Mode
Neonatal Mode
SYS: 40 mmHg ~ 260 mmHg
DIA: 20 mmHg ~ 200 mmHg
MAP: 26 mmHg– 220 mmHg
SYS: 40 mmHg– 230 mmHg
DIA: 20 mmHg– 160 mmHg
MAP: 26 mmHg– 183 mmHg
SYS: 40 mmHg – 130 mmHg
DIA: 20 mmHg– 100 mmHg
MAP: 26 mmHg – 110 mmHg
SYS, DIA, MAP
1 mmHg
±5 mmHg
8 mmHg
130 s
90 s
75 s
< 300 mmHg
< 150 mmHg
Default: 160 mmHg
Range:120/140/150/160/180/200/220/240/260/280 mmHg
Default: 140 mmHg
Range:
80/100/120/140/150/160/180/200/220/250 mmHg
Default: 90 mmHg
Range: 60/70/80/90/100/120/140 mmHg
- 254 -
Product Specification