EDAN iM50 User Manual page 79

Patient monitor
Hide thumbs Also See for iM50:
Table of Contents

Advertisement

Patient Monitor User Manual
Message
CO
Hardware Error
2
CO
Software Error
2
CO
Motor Error
2
CO
Factory
2
Calibration Data Lost
CO
Air
Pressure
2
Overrange
CO
No Sample Line
2
CO
No Adapter
2
CO
Occlusion
2
CO
Noisy Signal
2
Cause
Hardware
error
of
module
Software
error
of
module
Malfunction in the CO
motor.
Ex-factory calibration data
of CO
module is lost.
2
The
barometric
exceeds
the
working barometric pressure
range.
No
sampling
tube
connected.
No adapter connected
CO
module sample line
2
occluded.
The CO
signal is interfered
2
by
ambient
electromagnetic interference
Alarm Level
CO
2
High
CO
2
Medium
2
High
Medium
pressure
specified
High
is
Medium
Medium
High
or
Low
- 67 -
Alarm Information
Action Taken
Stop
measuring
function
of
CO
module, and notify
biomedical engineer
or
Manufacturer's
service staff.
Make sure the CO
module
is
used
within the specified
barometric pressure
range.
Connect
the
sampling
tube
correctly.
Connect the adapter
correctly.
Replace the sampling
line.
Check
interference
sources around the
device.
2
2