When the monitor is working, do not insert or unplug G7 module (with O
inaccurate O
measurement.
2
14
Do not place the module in any position that might cause it to fall on the patient.
15
Do not immerse sampling lines in liquid.
16
It is recommended to use two agents maximally at the same time.
1
Do not operate the module outside the specified operating temperature environment.
2
The sidestream AG module with O
3
The module should be securely mounted in order to avoid the risk of damage to the
module. The module should be handle with care and it cannot be fallen or thrown.

19.2.2 Safety Information for Masimo ISA Analyzer

1
The ISA sidestream gas analyzer is intended for use by authorized healthcare
professionals only.
2
Carefully route the sampling line to reduce the risk of patient entanglement or
strangulation.
3
Do not lift the ISA sidestream gas analyzer by the sampling line as it could disconnect
from the ISA, causing the ISA sidestream gas analyzer to fall on the patient.
4
Dispose Nomoline Family sampling lines in accordance with local regulations for
biohazardous waste.
5
Use only airway T-adapters with the sampling point in the center of the adapter.
6
Do only use sample lines intended for anesthetic agents if N
agents are being used.
7
Do not use the T-adapter with infants or neonates, as this adds 7 ml dead space to the
patient circuit.
8
Do not use the ISA gas analyzer with metered-dose inhalers or nebulized medications
as this may clog the bacteria filter.
9
Since a successful zeroing requires the presence of ambient air (21% O
CO
), ensure that the ISA is placed in a well ventilated place. Avoid breathing near the
2
ISA sidestream gas analyzer before or during the zeroing procedure.
10
Never sterilize or immerse the ISA sideateam gas analyzer in liquid.
WARNING
CAUTION
is fragile and should be handled with care.
2
WARNING
- 168 -
Monitoring AG
) to prevent
2
O and/or anesthetic
2
and 0%
2