EDAN iM50 User Manual page 261

Patient monitor
Hide thumbs Also See for iM50:
Table of Contents

Advertisement

Patient Monitor User Manual
Accuracy of Heart Rate
Meter and Response to
Irregular Rhythm
Time to Alarm for Heart
Rate alarm conditions
Arrhythmia
analyses
(Advanced)
Arrhythmia
analyses
(Basic)
12-Lead
Synchronization Analysis
Complied with IEC 60601-2-27: 2011, Sect. 201.7.9.2.9.101 b)
4), the HR value after 20 seconds of stabilization is displayed as
follows:
Ventricular bigeminy: 80 bpm±1 bpm
Slow alternating ventricular bigeminy: 60 bpm±1 bpm
Rapid alternating ventricular bigeminy: 120 bpm±1 bpm
Bidirectional systoles: 91 bpm±1 bpm
Asystole alarm: ≤ 10 s
HR low alarm: ≤ 10 s
HR high alarm: ≤ 10 s
Asystole
Vent Rhythm
Tachy
Irr Rhythm
Pacer not Pacing
VEB
IPVC
Pauses/min High
PAC Bigeminy
ExtremeBrady
Sustain VT
ASYSTOLE
VT > 2
VENT
TACHY
IRR
PNP
ECG
Average parameters of heart beat
Heart rate (bpm)
Time limit of P wave (ms)
PR interval (ms)
QRS interval (ms)
V-Fib/V-Tach
PVC Bigeminy
R on T
Brady
Vent Brady
Run PVCs
Non-Sustain VT
Pause
PVCs High
PAC Trigeminy
ExtremeTachy
VFIB/VTAC
BIGEMINY
R on T
BRADY
VBRADY
- 249 -
Product Specification
Couplet
PVC Trigeminy
PVC
Missed Beat
Pacer not Capture
Acc. Vent Rhythm
Multiform PVCs
Afib
Low Voltage(Limb)
Wide QRS Tachy
V-Tach
COUPLET
TRIGEMINY
PVC
MISSED BEATS
PNC