Patient Monitor User Manual
secures the power cable.
7
Power supply interface: for connecting AC power cable.
Network interface. It connects the monitor to the central monitoring system (also
named as MFM-CMS), CMS-LITE or gateway via standard network cable, which
8
enables MFM-CMS or gateway to achieve bidirectional communication with the
monitor.
Multifunctional port: Nurse call port/analog output/defibrillator synchronization.
Nurse call port: it connects the monitor to the hospital's nurse call system. Alarms
indications are alerted through the nurse call system if configured to do so.
9
Analog output: the monitor outputs the waveform through the port.
Defibrillator
synchronization signal through the port.
USB interfaces: supports USB 2.0 output. It connects approved USB devices, for
10
example, USB flash disk, barcode scanner, mouse and keyboard.
11
VGA output: it enables the VGA video output.
1
2
3
4
5
6
Handle: for lifting or moving the monitor, which is convenient, comfortable and
1
fast.
2
Label
3
VGA output: it enables the VGA video output.
synchronization:
the
iM60/iM70
- 20 -
monitor
outputs
7
8
9
Basic Operation
the
defibrillator
12
11
10