EDAN iM50 User Manual page 70

Patient monitor
Hide thumbs Also See for iM50:
Table of Contents

Advertisement

Patient Monitor User Manual
Message
SpO
Low Perfusion
2
(EDAN SpO
module)
2
SpO
Noisy Signal
2
(EDAN SpO
module)
2
SpO
Light
2
Interference
(EDAN SpO
module)
2
NIBP
NIBP Comm Fail
NIBP Leak
Cause
The pulse signal is too
weak or the perfusion of
the measurement site is too
low. The SpO
value and
2
PR
value
might
inaccurate then.
There is interference with
SpO
measurement signals
2
due to patient movement,
ambient
light,
electrical
interference or else.
Ambient light around the
sensor is too strong.
NIBP module failure or
communication failure
NIBP pump, valve, cuff or
tube has a leakage.
Alarm Level
Low
be
Low
Low
High
Low
- 58 -
Alarm Information
Action Taken
Reconnect the SpO
sensor and change
the
measurement
site.
If
problem
exists, please notify
biomedical engineer
or
manufacturer's
service staff.
Check the condition
of patient and avoid
patient
movement;
make sure the cable
is well connected.
Reduce interference
of the ambient light
and avoid sensor's
exposure to strong
light.
Stop
using
measuring
function
of NIBP module, and
notify
biomedical
engineer
or
manufacturer's
service staff.
Check
the
connections and the
wrapped cuff to see
whether they are all
prepared well.
If failure persists,
please
notify
biomedical engineer
or
manufacturer's
service staff.