EDAN iM50 User Manual page 120

Patient monitor
Hide thumbs Also See for iM50:
Table of Contents

Advertisement

Patient Monitor User Manual
ARR Alarms
Basic ARR
ASYSTOLE
VFIB/VTAC
VT > 2
COUPLET
BIGEMINY
TRIGEMINY
R ON T
PVC
TACHY
BRADY
MISSED BEATS
IRR
PNC
PNP
VBRADY
VENT
NOTE: Arrhythmia monitoring is intended for use with adult patients only and is not clinically
validated for use with neonatal and pediatric patients. For this reason, the recommended
setting for arrhythmia monitoring in neonatal and pediatric modes is Off.
Selecting an ECG lead for Arrhythmia:
In arrhythmia monitoring, it is important to select the appropriate lead.
For non-paced patients, the guidelines are:
No QRS is detected for 4 consecutive seconds
4 consecutive seconds' fibrillation wave occurs, or each RR interval for 5
consecutive ventricular beats is less than 600 ms.
3 ≤ the number of consecutive PVCs < 5
2 consecutive PVCs
A dominant rhythm of N, V, N, V (N = supraventricular beat, V =
ventricular beat) was detected.
A dominant rhythm of N, N, V, N, N,V
A type of single PVC under the condition that HR < 100,R-R interval is
less than 1/3 the average interval, followed by a compensating pause of
1.25X the average R-R interval (the next R wave advances onto the
previous T wave).
Single PVC detected in normal heartbeats.
Adult: RR interval for 5 consecutive QRS complex ≤ 0.5 s.
Pediatric/neonatal: RR interval for 5 consecutive QRS complex ≤ 0.375 s.
Adult: RR interval for 5 consecutive QRS complex ≥ 1.5 s.
Pediatric/neonatal: RR interval for 5 consecutive QRS complex ≥ 1 s.
If HR < 120 bpm, no beats are detected for 1.75 times average RR
interval; or if HR ≥ 120 bpm, no beats are detected for one second.
Consistently irregular heart rhythm
PACER NOT CAPTURE: no QRS complex detected in 300 ms after a
pace pulse.
PACER NOT PACED: no pace pulse detected in 1.75 times RR interval
after a QRS complex.
VENTRICULAR BRADYCARDIA: Each RR interval for 5 consecutive
ventricular beats > 1000 ms.
VENTRICULAR RHYTHM: Each RR interval for 5 consecutive
ventricular beats ranges from 600 ms to 1000 ms.
Occurring Condition
- 108 -
Monitoring ECG