Patient Monitor User Manual
49
The monitor is suitable for use in the presence of electrosurgery. When the monitor is
used with HF surgical equipment, user (doctor or nurse) should be cautious about
patient safety.
50
Make sure networking function is used in a secure network environment.
1
Electromagnetic Interference - Ensure that the environment in which the patient
monitor is installed is not subject to any sources of strong electromagnetic
interference, such as radio transmitters, mobile telephones, microwaves, etc.
2
Keep the environment clean. Avoid vibration. Keep it far away from corrosive
medicine, dust area, high temperature and humid environment.
3
Do not immerse transducers in liquid. When using solutions, use sterile wipes to avoid
pouring fluids directly on the transducer.
4
Do not use autoclave or gas to sterilize the monitor, recorder or any accessories.
5
The device and reusable accessories may be sent back to the manufacturer for
recycling or proper disposal after their useful lives.
6
Disposable devices are intended for single use only. They should not be reused as
performance could degrade or contamination could occur.
7
Remove a battery whose life cycle has expired from the monitor immediately.
8
Avoid liquid splashing on the device.
9
To ensure patient safety, use only parts and accessories manufactured or
recommended by EDAN.
10
Before connecting the monitor to the AC power, make sure the voltage and the power
frequency are consistent with the requirements indicated on the device label or in this
user manual.
11
Protect the device against mechanical damage resulting from falls, impacts, and
vibration.
12
Do not touch the touch screen with a sharp object.
13
A ventilated environment is required for monitor installation. Do not block up the
ventilation grille at the back of the device.
14
The device must be connected to the ground to avoid the signal interference.
15
Poor connection might be caused by frequently plugging and unplugging the power
cord. Check the power cord regularly and replace it in time.
16
To protect eyes from damage, don't look directly at supplementary light for long time.
17
Extremely frequent clicking on the screen within a short period of time may result in
temporary discontinuous data.
WARNING
CAUTION
- 6 -