Patient Monitor User Manual
Wi-Fi Network Stability
Distinct Vision Distance
◆ ESU equipment
The wireless network function of all monitors
works
normally
requirements:
■ Total delay time for data transmission from the
monitors to MFM-CMS: ≤ 2 s.
■ Total delay time of data transmission from one
monitor to other monitors: ≤ 2 seconds.
■ Effective time of alarm reset configured on
another monitor ≤ 2 s.
■ Effective time for monitor-related settings
configured on the MFM-CMS: ≤ 2 s.
■ No communication loss between all the monitors.
When the following conditions are present,
■ Quantity of the monitors supported by a single
AP: ≤ 8.
■ Each monitor can communicate with MFM-CMS.
■ Each monitor supports bed view function, which
allows users to view its information from another
bed or view other bed's information from its screen.
■ The AP signal strength of the monitor should be
stronger than -65 dBm.
The following requirements must be met:
■ Within 24 hours, the time percentage of failing to
transmit data from any monitor to the MFM-CMS
does not exceed 0.1%. When the connected 8
monitors roam for 30 times, the time percentage of
failing to transmit data from any monitor to the
MFM-CMS does not exceed 0.1%.
The distinct vision distance between the monitor
and the AP: ≥ 50 meters.
- 243 -