Patient Monitor User Manual
• Quaternary Ammonium Chloride Wipe
• 0.5% Sodium Hypochlorite/Water Solution
• Accelerated Hydrogen Peroxide
Maintenance
Once every year it is recommended to perform maintenance on the module. A NomoLine ISA
CO
Maintenance Kit containing all required components and instructions to perform
2
maintenance procedures is available through www.masimo.com.
17.3.2.2 Removing Exhaust Gases from the System
Do not connect the exhaust tube to the ventilator circuit, connect the outlet to a
scavenging system, cross infection can occur if sampling gas is returned to the breathing
system. When using the sidestream CO
have recently received anesthetics, please avoid exposing medical staff to anesthetics.
Use an exhaust tube to remove the sample gas to a scavenging system. Attach it to the sidestream
sensor at the outlet connector.
17.3.3 Mainstream CO
For Respronics Module:
NOTE:
You must perform a zero calibration as described in this procedure each time you use a
new airway adapter.
17.3.3.1 Measurement Steps
1
Attach the sensor connector to the CO
2
Wait two minutes, allowing the sensor to reach its operating temperature and a stable
thermal condition.
3
Choose the appropriate airway adapter and connect it to the sensor head. The airway adapter
clicks into place when seated correctly.
WARNING
measurement on patients who are receiving or
2
Module
2
connector on the monitor.
2
- 157 -
Monitoring CO
2