EDAN iM50 User Manual page 299

Patient monitor
Hide thumbs Also See for iM50:
Table of Contents

Advertisement

Patient Monitor User Manual
Immunity test
Electrostatic
discharge (ESD)
IEC/EN 61000-4-2
Electrical fast
transient/burst
IEC/EN 61000-4-4
Surge
IEC/EN 61000-4-5
Power frequency
(50/60Hz)
magnetic field
IEC/EN 61000-4-8
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC/EN
61000-4-11
NOTE
U
is the a.c. mains voltage prior to application of the test level.
T
IEC/EN 60601 test level
± 8 kV contact
± 15 kV air
± 2 kV for power supply
lines
± 1 kV for line to line
± 2 kV for line to ground
30 A/m
0 % U
0.5 cycle
T;
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
0 % U
; 1 cycle
T
and
70 % U
; 25/30 cycles )
T
Single phase: at 0°
0 % U
; 250/300 cycle
T
Compliance level
± 8 kV contact
± 15 kV air
± 2kV for power
supply lines
± 1 kV for line to
line
± 2 kV for line to
ground
30 A/m
0 % U
0.5 cycle
T;
At 0°, 45°, 90°,
135°, 180°, 225°,
270° and 315°
0 % U
; 1 cycle
T
and
70 % U
; 25/30
T
cycles )
Single phase: at 0°
0 % U
; 250/300
T
cycle
- 287 -
EMC Information
Electromagnetic
environment - guidance
Floors should be wood,
concrete or ceramic tile.
If floor are covered
with synthetic material,
the relative humidity
should be at least 30%.
Mains power quality
should be that of a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment.
Power
frequency
magnetic fields should
be
at
levels
characteristic
of
typical location in a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment. If
the user of iM50, iM60,
iM70, iM80, M50 and
M80 requires continued
operation during power
mains interruptions, it is
recommended
iM50,
iM60,
iM70,
iM80, M50 and M80 be
powered
from
uninterruptible
power
supply or a battery.
a
that
an