Satseconds "Safety Net; Setting Satseconds Duration - EDAN iM50 User Manual

Patient monitor
Hide thumbs Also See for iM50:
Table of Contents

Advertisement

Patient Monitor User Manual
After approximately 10.7 seconds, a SatSeconds alarm will be triggered, because the limit of 50
SatSeconds has been exceeded. See arrow (↑) in the following figure.
Saturation levels may fluctuate rather than remaining steady for a period of several seconds.
Often, the patient SpO
range several times. During such fluctuation, the monitor integrates the number of SpO
both positive and negative, until either the SatSeconds limit is reached, or the patient SpO
returns within a normal range and remains there.

11.11.2 SatSeconds "Safety Net"

The SatSeconds "Safety Net" is for patients whose saturation makes frequent excursions below or
above the SpO
limit but does not remain in violation long enough for the SatSeconds limit to be
2
reached. If three or more SpO
will be triggered even if the SatSeconds limit has not been reached.

11.11.3 Setting SatSeconds Duration

You can set SatSeconds to Off or to the duration among 10 , 25 , 50 and 100 . To configure the
SatSeconds settings, enter the SpO
the SatSeconds list.
SpO
2
2
×
4
×
6
×
Total SatSeconds
Alarm Response with SatSeconds
may fluctuate above and below the alarm limit, re-entering the non-alarm
2
alarm limit violations occur within a 60-second period, an alarm
2
Setup menu and select the desired SatSeconds setting from
2
Seconds
2
=
3
=
6
=
=
- 125 -
Monitoring SpO
SatSeconds
4
12
36
52
2
points,
2
2