Patient Monitor User Manual
Interfering Gas Effects:
Gas
Xenon/Helium/Metered dose
inhaler propellants
Ethanol/Isopropanol/Acetone 0.1%
Methane
Masimo ISA analyzer
Module Type
Measurement
Parameters
Measurement
Principle
Dimensions (W × D ×
H)
Weight
According
to
degree of protection
against
harmful
ingress of water
Ambient CO
2
Mechanical
robustness
Recovery time after
defibrillator test
Drift of measurement
accuracy
Gas Level (%)
Not applicable
1%
ISA AX+
Displaying the concentration of CO
anaesthesia agent and identifying the anaesthesia agent
automatically (built-in module)
ISA OR+
Displaying the concentration of CO
anaesthesia agent and identifying the anaesthesia agent
automatically (built-in module)
CO
, N
O , O
2
2
2
Sevoflurane (SEV) , Desflurane (DES), AwRR, MAC
CO
, N
O, Anaesthesia Agent: infra-red absorption characteristic;
2
2
O
: Paramagnetic method
2
ISA AX+: 33 mm × 78 mm × 49 mm (1.3" × 3.1" ×1.9")
ISA OR+: 49 mm × 90 mm × 100 mm (1.9" × 3.5" ×3.9")
Exclduing cable, tubing and Nomoline.
ISA AX+: < 130 g (cable included)
ISA OR+: < 420 g (cable included)
the
IPX4
≤ 800 ppm (0.08 vol%)
Meets the shock and vibration requirements EN ISO 80601-2-55:2011
clause 201.15.3.5.101.1
Unaffected
No drift
, Halothane (HAL), Isoflurane (ISO), Enflurane (ENF),
- 272 -
Product Specification
Quantitative Effect/Comments
Not applicable
None
None
, N
2
, O
, N
2
2
O, and two
2
O, and two
2