EDAN iM50 User Manual page 69

Patient monitor
Hide thumbs Also See for iM50:
Table of Contents

Advertisement

Patient Monitor User Manual
Message
RESP Noise
RR Exceed
SpO
2
SpO
Sensor Off
2
SpO
Comm Fail
2
SpO
Sensor Err
2
SpO
No Sensor
2
Cause
RR cannot be measured due
to patient movement.
RR measuring value is out
of the measure range.
SpO
sensor
may
2
disconnected
from
patient measuring site.
SpO
module failure or
2
communication failure
Malfunction in the SpO
sensor or in the extension
cable.
No
SpO
sensor
2
connected to the monitor.
Alarm Level
Low
Medium
be
the
Low
High
2
Low
was
Low
- 57 -
Alarm Information
Action Taken
Check whether the
RESP leads are well
connected. Keep the
patient
calm
for
better monitoring.
Check
whether
interference to the
respiratory
signal
exists.
And
check
whether the patient is
breathing
normally;
breathing too rapidly
or too slowly may
endanger
patient's
life.
Make sure the sensor
is well connected to
the patient's finger
or other parts. Make
sure the monitor and
cables
are
well
connected.
Stop
using
measuring
function
of SpO
module, and
2
notify
biomedical
engineer
or
manufacturer's
service staff.
Replace the SpO
sensor
or
the
extension cable.
Make
sure
the
monitor and sensor
are well connected,
reconnect the sensor.
2

Advertisement

Table of Contents
loading

This manual is also suitable for:

Im60Im70Im80M50M80