Sensor Type
Refresh Time

6.1.6 Edan Transducer Socket Board (wired model only)

The EDAN transducer socket board locates on the left of the lower case, i.e. the left of the
monitor. It communicates between the main control board and external transducer accessories. Its
function is to isolate the signal of accessories, power supply and ground, and transfers collected
SpO2 and TEMP signals to iM X12 parameter board.
Here is the principle block diagram:
Socket to
connect with
lower case
interface
board
Figure 6-5 EDAN Transducer Socket Board Principle Block Diagram
Note:
The US, TOCO&MECG and DECG&IUP transducer can be plugged in and out at any of the 4
transducer interface.

6.1.7 NELL Transducer Socket Board (wired model only)

The NELL transducer socket board locates at left side of the lower case, i.e. the left of the
monitor. The NELL transducer socket board communicates between the main control board and
external transducer accessories. Its function is to isolate the signal of accessories, power supply
and ground, and transfers collected SpO2 and TEMP signals to iM X12 parameter board.
Here is the principle block diagram:
YSI-10K
Every 1 ~ 2 second(s)
Communicate
signal*4
MARK、EXT
Isolation
+5V_I
N
Circuit
+12V
+12V time-
+12V_I
delay
N
circuit
Reflective
Interrupter*4
Reflective
Interrupter*2
Supply power
+5V5
for transducer
accessories
LDO
- 28 -
Principle Introduction
Transducer Interface*4
(support any fetal
transducer at any
interface)
MARK_ISO、
EXT_ISO
+5V_IS
O