Message
NIBP Pulse Signal
Weak
(Omron NIBP module)
TEMP
TEMP T1 Sensor Off
TEMP T2 Sensor Off
Excessive T1
Excessive T2
TEMP Comm Fail
T1 Calibration Failed
T2 Calibration Failed
IBP
YY Sensor Off (YY
stands for the IBP label
name: Art, PA, CVP,
RAP, LAP, ICP, P1
and P2)
Cause
Pulse is too weak, and the
detected signal is too weak.
Temperature cable of TEMP
channe1
may
disconnected
from
monitor.
Temperature cable of TEMP
channe2
may
disconnected
from
monitor.
TEMP1 measuring value is
beyond measuring range.
TEMP2 measuring value is
beyond measuring range.
TEMP module failure or
communication failure.
T1 calibration failed.
T2 calibration failed
IBP sensor falls off.
Alarm Level
Low
be
Low
the
be
Low
the
High
High
High
High
High
Medium
- 61 -
Alarm Information
Action Taken
Check if the cuff
leak
or
cuff
is
properly wrapped.
Make sure that the
cable
is
properly
connected
Make sure that the
cable
is
properly
connected.
Check
sensor
connection
and
patient condition
Check
sensor
connection
and
patient condition
Stop
measuring
function of TEMP
module, and notify
biomedical engineer
or
Manufacturer's
service staff.
Please check whether
the module works
properly.
Please check whether
the module works
properly.
Check
the
sensor
connection
and
reconnect the sensor.