EDAN iM50 User Manual page 211

Patient monitor
Hide thumbs Also See for iM50:
Table of Contents

Advertisement

Patient Monitor User Manual
RESP (rpm)
SpO
(%)
2
TEMP ( ℃ )
SYS (mmHg)
HR (bpm)
Consciousness
Oxygen
In MEWS interface, select Criteria to check score criteria as following:
HR (bpm)
SYS (mmhg)
RESP (rpm)
TEMP ( ° C)
Consciousness
Note: (1) HR=40, value is 2; (2) SYS=70, value is 3.
The relationship between consciousness level and its display result is as below:
3
2
≤8
≤91
92~93
≤35.0
35.1~36.0 36.1~38.0 38.1~39.0
≤90
91~100
101~110
≤40
Yes
3
2
<40
41~50
<70
71~80
81~100
<9
<35.0
Consciousness
Sober
Responsive to Voice
Responsive to Pain
Unresponsive
Value
1
0
9~11
12~20
94~95
≥96
111~219
41~50
51~90
A
No
Value
1
0
51~100
101~199
9~14
35.0~38.4
A
Displayed Result
- 199 -
Warning-Score System
1
2
21~24
≥39.1
91~110
111~130
1
2
101~110
111~129
≥200
15~20
21~29
≥38.5
V
P
A
V
P
U
3
≥25
≥220
≥131
V/P/U
3
≥130
≥30
U