EDAN iM50 User Manual page 86

Patient monitor
Hide thumbs Also See for iM50:
Table of Contents

Advertisement

Patient Monitor User Manual
Message
AG Agent Estimated
(Dräger AG module)
AG Interference
Warm-up over
Measure over
Measure time out
Printer Busy
No Default Printer
Incomplete parameter input,
unable to score
No WIFI module detected
SpO
Noisy Signal
2
(Nellcor SpO
module)
2
Recorder Setup Needed
NIBP Simul
The space in U disk is less
than 300 M. Please clean it up.
Please input user password
first.
Attention!
information included in the
data.
More than five consecutive
password errors
Cause
The AG module cannot identify the present agent(s) but
only give an estimation of one of the present agents. The
reason is the presence of either a mixture or too many
anesthetic.
AG module is interfered by external device when working.
The monitor displays this message after taking the sensor
out of the bracket and warm-up is over.
After the Predict measuring is over, the data and message
display on the interface.
No measuring result after the module entering Predict
state for 30 s.
The monitor is performing a print job.
No default printer has been set.
In Warning-Score System interface, parameters are not
completely input.
No Wi-Fi module is detected.
There is interference with SpO
to patient movement, ambient light, electrical interference
or else.
The user presses the RECORD button or shortcut key
Record when Recorder is not configured.
NIBP Simul function is turned on.
The remaining space of U disk is less than 300 M.
Private
When user exports data from the internal storage device.
Continuously enter the wrong password for more than 5
times.
- 74 -
Alarm Information
measurement signals due
2

Advertisement

Table of Contents
loading