Mainstream - EDAN iM50 User Manual

Patient monitor
Hide thumbs Also See for iM50:
Table of Contents

Advertisement

Patient Monitor User Manual
1
☆ Accuracy
Rise Time (t10...90)
Time to Specified Accuracy
Automatic Detection
Primary Gas
Secondary Gas
☆ Respiratory Rate
☆ Range
☆ Accuracy
Resolution
Note 1: In accordance to ISO 21647:2004 and ISO 80601 ‑ 2 ‑ 55:2011, for respiratory rates from
1
0...60
/min with I:E ratio of 1:1.
Note 2: Duration from power on at 10 °C module temperature to transmission of measurements
with unspecified accuracy
Note 3: Duration from power on at 10 °C module temperature to transmission of measurements
with specified accuracy
Note 4: With Dräger sample line and water trap
A.14.2 Mainstream
Masimo IRMA module
Module Type
Measurement
Parameters
Measurement
Principle
± (0.2 Vol% + 15 % rel.)
4
< 450 ms
3
< 450 s
At the latest at 0.3 Vol%
At the latest at 0.4 Vol%
With a Desflurane concentration greater than 4 Vol%,
mixture detection occurs at the latest when the concentration
of the second anesthetic gas rises above 10% of the
Desflurane concentration.
0/min to 100/min (Respiratory rate is derived from the CO
value)
0/min to 60/min: ± 1 /min
> 60 /min: not specified
1 /min
IRMA AX+
Displaying the concentration of CO
anaesthesia agent and indentifying two anaesthesia
agent
CO
, N
O, HAL, Isoflurane (ISO), Enflurane (ENF),
2
2
Sevoflurane (SEV), Desflurane (DES), AwRR, MAC
CO
, N
O, anaesthesia agent: infra-red absorption characteristic
2
2
- 278 -
Product Specification
2
, N
O and two
2
2