EDAN iM50 User Manual page 257

Patient monitor
Hide thumbs Also See for iM50:
Table of Contents

Advertisement

Patient Monitor User Manual
☆ Accuracy of Signal
Reproduction
Electrode
Potential Tolerance
Auxiliary Current (Leads
off detection)
☆ Recovery Time After
Defibrillation
Leakage Current of Patient < 10 μA
Scale Signal
☆ System Noise
☆ Multichannel Crosstalk
☆ Frequency and Impulse
Response
Sampling Frequency
Sampling Channel Switch
Time
A/D Precision
☆ ESU Protection
Electrosurgical
Interference Suppression
Minimum Input Slew Rate
(Lead II)
☆ Baseline Reset Time
An error of ≤ ±20 % of the nominal value of the output or ±100
μV, whichever is greater.
The total error and frequency response comply with IEC
60601-2-27: 2011, Sect. 201.12.1.101.1.
±800 mV
Offset
Active electrode: < 100 nA
Reference electrode: < 900 nA
< 5 s (measured without electrodes as IEC60601-2-27:2011, Sect.
201.8.5.5.1 requires.)
1 mV PP, accuracy is ±5%
< 30 μVPP
≤ 5% of the input signal
Complied with IEC 60601-2-27: 2011, Sect. 201.12.1.101.5.
Frequency response:
Input a 5 Hz, 1 mV sine wave signal, and the output signal
amplitude remains within the range of 71 % to 110 % at 0.67 Hz
and 40 Hz.
Input a 1 Hz, 1.5 mV 200 ms triangular wave input signal, and
the output shall be within 11.25 mm~15 mm.
Impulse response:
Displacement value: ≤ 0.1 mV
Slope: ≤ 0.3 mV/s following the end of the pulse.
Complied with IEC 60601-2-27: 2011, Sect. 201.12.1.101.8.
1000 Hz
< 80 μS
24 Bits (Minimum resolution: 0.077uV/LSB)
Cut mode: 300 W
Coagulation mode: 100 W
Restore time: ≤ 10 s
Test according to ANSI/AAMI EC13:2002, Sect. 5.2.9.14.
Complied with ANSI/AAMI EC13:2002, Sect. 4.2.9.14.
> 2.5 V/s
< 3 s
- 245 -
Product Specification