Patient Monitor User Manual
Rise
time
(@
sml/min sample flow)
Primary Agent
Threshold
Secondary Agent
Threshold
Agent Identification
Time
Total system response
time
Work Mode
Data Update Period
Measurement Range
Resolution
Gas
Accuracy-
CO
2
Standard
Conditions
N
O
2
HAL, ENF, ISO
SEV
DES
CO
≤ 300 ms
2
N
O, O
, ENF, ISO , SEV, DES ≤ 400 ms
50
2
2
HAL ≤ 500 ms
(Measured according to EN ISO 80601-2-55)
0.15 vol%. When an agent is identified, concentrations will be
reported even below 0.15 vol% as long as apnea is not detected.
0.2 vol% + 10% of total agent concentration
< 20 seconds (typically < 10 seconds)
< 4 seconds (with 2 m Nomoline Airway Adapter Set sampling line)
Measure
1 s
CO
: 0 to 25 vol%
2
O
: 0 to 100 vol%
2
N
O: 0 to 82vol%
2
HAL, ENF, ISO, SEV, DES: 0 to 25 vol%
AwRR: 0 rpm to 150 rpm
CO
: 0.1%
2
HAL, ENF, ISO, SEV, DES: 0.1%
N
O: 1%
2
O
:1%
2
AwRR: 1 rpm
Range
0 to 15 vol%
15 to 25 vol%
0 to 82 vol%
0 to 8 vol %
8 to 25 vol %
0 to 10 vol %
10 to 25 vol %
0 to 22 vol %
22 to 25 vol %
Accuracy
±(0.2 vol% + 2% of reading)
Unspecified
±(2 vol% + 2% of reading)
±(0.15 vol% + 5% of reading)
Unspecified
±(0.15 vol% + 5% of reading)
Unspecified
±(0.15 vol% + 5% of reading)
Unspecified
- 274 -
Product Specification