Troubleshooting
Resetting the
Analyzer
Thermo Scientific Orion 2117HL High Level Chloride Analyzer User Guide
The Thermo Scientific Orion 2117HL chloride analyzer allows the user to
reset the analyzer through the setup mode or by a hard reset. Resetting the
analyzer will restore all settings in the setup mode to factory default values.
Warning: Resetting the analyzer will erase all data logs including
calibration, and password protection settings for setup and calibration.
The analyzer will reset setup parameters to factory default settings.
Operator Action
Press
in the measure mode
to enter the setup mode
/
Press
until RSET
appears in the marquee
Press
to reset the analyzer
Press
Press
Press
When the reset is complete, the
system will return to the measure
mode. The operator will need
to re-enter the setup mode to
continue programming the setup
parameters. Press
to return
to the setup mode.
Scrolling
Main
Marquee
Display
Last menu used
SEL
in setup mode
SCrn
RSET
SEL
(flashing)
SCrn
PRESS TEST
rSEt
VIEW ENTER TO
?
RESET
PRESS TEST
rSEt
VIEW ENTER TO
?
RESET
Notes
SETUP appears
as mode
This command
resets all
previously set
parameters to
factory default
values. Use this
command only to
set the analyzer
to original
factory setup
values.