Thermo Scientific Orion 2117HL User Manual page 57

High level chloride analyzer
Table of Contents

Advertisement

Analyzer Operation
Default
Operator Action
TEST (cont'd)
To Test rLY:
4-20
Press
4-20 and rLY settings and select rLY
Press
Actual
Press
rLY1 status
setting OFF or On
Press
move to the rLY2 test
Actual
Press
rLY2 status
setting OFF or On
Press
move to the rLY3 test
Actual
Press
rLY3 status
setting OFF or On
Press
return to the main setup mode
Press
specific menu options in the setup mode
Press
Thermo Scientific Orion 2117HL High Level Chloride Analyzer User Guide
/
to scroll between
to test relay outputs
/
to set the rLY1
to accept the rLY1 test and
/
to set the rLY2
to accept the rLY2 test and
/
to set the rLY3
to accept the rLY3 test and
to return to the channel
to scroll to the next menu
Scrolling Marquee
SELECT mA OR RELAY
USE ARROWS TO TOGGLE
RELAY 1
USE ARROWS TO TOGGLE
RELAY 2
USE ARROWS TO TOGGLE
RELAY 3
CH1 or CH2
(flashing)
TEST
(flashing)
Main Display
Notes
tESt
rLY
(flashing)
rLY1
Relay contact
according to screen
OFF or On
indication
rLY2
Relay contact
according to screen
OFF or On
indication
rLY3
Relay contact
according to screen
OFF or On
indication
SEL
Display for two
channel analyzer only
SCrn
Action for two
channel analyzer only
SEL
SEL CH1 or SEL CH2
on main display of
SCrn
two channel analyzer
III-31