Thermo Scientific Orion 2117HL User Manual page 98

High level chloride analyzer
Table of Contents

Advertisement

Troubleshooting
Displayed
Description
Event Code
System Event/Error Codes - Yellow LED
E004
DEFAULT VALUES
E021
HARDWARE FAULT
E101
CH1 TEMP OUT OF RANGE
E102
CH2 TEMP OUT OF RANGE
E111
CH1 CAL DUE
E112
CH2 CAL DUE
E127
CH1 SOUT mA OVER RANGE
E128
CH1 SOUT mA UNDER RANGE Measured sensor value below Sout low
E129
CH1 TOUT mA OVER RANGE
E130
CH1 TOUT mA UNDER RANGE
E131
CH2 SOUT mA OVER RANGE
E132
CH2 SOUT mA UNDER RANGE Measured sensor value below Sout low
E133
CH2 TOUT mA OVER RANGE
E134
CH2 TOUT mA UNDER RANGE
Thermo Scientific Orion 2117HL High Level Chloride Analyzer User Guide
Cause
Analyzer has been reset or is new out
of box
System error
Temperature sensor on channel 1 out of
range (5 °C to 45 °C) for ISE
Temperature sensor on channel 2 out of
range (5 °C to 45 °C) for ISE
User calibration required on channel 1.
No calibration has been performed in
user specified time interval.
User calibration required on channel 2.
No calibration has been performed in
user specified time interval.
Measured sensor value above Sout high
limit value on channel 1
limit value on channel 1
Measured temperature value above Tout
high limit value on channel 1
Measured temperature value below Tout
low limit value on channel 1
Measured sensor value above Sout high
limit value on channel 2
limit value on channel 2
Measured temperature value above Tout
high limit value on channel 2
Measured temperature value below Tout
low limit value on channel 2
Troubleshooting
Calibrate the analyzer
Contact Technical Support 1-800-225-1480
Check ATC cable and probe connections. Replace if needed.
Check ATC cable and probe connections. Replace if needed.
Recalibrate the analyzer.
Recalibrate the analyzer.
Verify user selectable Sout high limit value in setup mode
Verify user selectable Sout low limit value in setup mode
Verify user selectable Tout high limit value in setup mode
Verify user selectable Tout low limit value in setup mode
Verify user selectable Sout high limit value in setup mode
Verify user selectable Sout low limit value in setup mode
Verify user selectable Tout high limit value in setup mode
Verify user selectable Tout low limit value in setup mode
VI-15