Thermo Scientific Orion 2117HL User Manual page 112

High level chloride analyzer
Table of Contents

Advertisement

Appendix
Thermo Scientific Orion 2117HL High Level Chloride Analyzer User Guide
Sample Conditions
Temperature
Total Alkalinity
Sulfite
Inlet Pressure
Flow Rate
Sample Inlet
Sample Drain
Grab Sampler
Display Features
Type
Size
Marquee
Middle Line
Lower Line
Text Display
Calibration
DKA Calibration
DKA Calibration Points
Offline Calibration
Offline Calibration Points
Pre-programmed Standard Values
Custom Calibration
Operator Input
Inputs (Potentiometric)
Ion Input
Reference Input
5 to 35 °C
Less than 50 ppm CaCO
3
Sulfite cannot be present in the sample
8 to 100 psig
40 mL/minute nominal set by pressure regulator
1/4" NPTF tube fitting
3/4" NPT male
Optional
Custom backlit LCD
54 x 76 mm (2 1/8 in x 3 in)
Temperature, operator prompts
Concentration, error codes
mV (selectable)
Scrolling marquee for English language instructions
on setup, calibration and diagnostics
Yes
3 points
Yes
1 point
Yes
Programmable for concentration and volume
DKA, operator programmable, offline
Chloride
Reference
A-4