Calibration
Step
Operator Action
10
Perform laboratory analysis on QC sample
and return with results.
11
Press
to resume the calibration.
12
If the calibration password is active, enter
the calibration password as described in
step 3.
Press
to accept the password.
13
14
Press
/
channel that will be calibrated and press
to begin the calibration.
15
Press
/
and press
to accept the setting.
16
Enter the concentration value of the QC
sample obtained from laboratory analysis:
Press
to move the decimal point
and set the value as ppm or ppb.
Press
/
Press
to move to the next digit.
Press
/
the remaining digits and press
move through the remaining digits.
Press
to save the value.
17
18
Press
.
Thermo Scientific Orion 2117HL High Level Chloride Analyzer User Guide
Scrolling Marquee
Actual temperature reading
ENTER PASSWORD
CH1 or CH2
(flashing)
CH1 or CH2
to select the
(flashing)
OFFL
to select OFFL
(flashing)
LAB ANALYSIS READY
INPUT CONC VALUE FROM LAB
PRESS ENTER TO CONTINUE OR
PRESS EXIT TO ABORT
to set the first digit.
to set the values of
to
NEW CALIBRATION Eo
PRESS ENTER TO ACCEPT OR
PRESS EXIT TO ABORT
Main Display
Notes
Actual concentration
System will return to
reading
measure mode and
operate as normal.
CAL and MEASURE
appear as mode.
HOLD and CAL appear
as mode.
PASS
# # # #
CAL
Displayed for two
channel analyzer
CHnL
CAL
Action required for
two channel analyzer
CHnL
CAL
tYPE
COnC
# # # . #
(change flashing digit)
Actual Eo value
The system
will display the
recalculated the E
The system will return
to the measure mode.
Only MEASURE
should appear
as mode.
0
IV-16