Thermo Scientific Orion 2117HL User Manual page 51

High level chloride analyzer
Table of Contents

Advertisement

Analyzer Operation
Default
Operator Action
mAMP
Press
4-20
Press
4-20 and 0-20
Press
advance to the next screen
LOg
Press
LOg and LIn
Press
advance to the next screen
1 ppb
Press
Press
Press
Press
the remaining digits and press
move through the remaining digits
Press
advance to the next screen
100 ppb
Press
Press
Press
Press
the remaining digits and press
move through the remaining digits
Press
advance to the next screen
Thermo Scientific Orion 2117HL High Level Chloride Analyzer User Guide
to set the analog output
/
to scroll between
to accept the setting and
/
to scroll between
to accept the setting and
to move the decimal point
/
to set the first digit
to move to the next digit
/
to set the values of
to
to accept the setting and
to move the decimal point
/
to set the first digit
to move to the next digit
/
to set the values of
to
to accept the setting and
Scrolling Marquee
mAMP
(flashing)
SELECT 0-20 OR 4-20
SELECT LOG OR
LINEAR OUTPUT
mA SENSOR OUTPUT
LOW VALUE
mA SENSOR OUTPUT
LOW VALUE
mA SENSOR OUTPUT
HIGH VALUE
mA SENSOR OUTPUT
HIGH VALUE
Main Display
SEL
SCrn
4-20 or 0-20
(flashing)
SOUt
LOg or LIn
(flashing)
LO
# # # . #
(first digit flashing)
LO
# # #. #
(change flashing digit)
HI
# # # . #
(first digit flashing)
HI
# # #. #
(change flashing digit)
Notes
SEL CH1 or SEL
CH2 on main
display of two
channel analyzer
III-25