Electrode Wiring Assignments - Thermo Scientific Orion 2117HL User Manual

High level chloride analyzer
Table of Contents

Advertisement

Analyzer Preparation
Electrode Wiring
Assignments
Thermo Scientific Orion 2117HL High Level Chloride Analyzer User Guide
Chloride Electrode
26
Sensing electrode
33
Shield
Reference Electrode
24
Reference electrode
32
Shield
2100TP Temperature Probe
16
Temperature ground /thermistor
17
Temperature drive /thermistor
19
Solution ground
2001TM Temperature Probe
16
Temperature ground /thermistor
17
Temperature drive /thermistor
18
Temperature sense
19
Solution ground
2001SC pH Electrode
24
Reference electrode
26
Sensing electrode
110250 ORP Electrode
19
Solution ground
24
Reference electrode
26
Sensing electrode
2002CC and 2002SS Conductivity Probes
6
Shield ground for conductivity
16
Temperature ground /thermistor
17
Temperature drive /thermistor
20
Conductivity drive +
21
Conductivity sense +
22
Conductivity sense -
23
Conductivity drive -
Note: Only reference the wiring configurations that are applicable to your
analyzer and electrodes.
Connect clear wire
Connect black wire
Connect clear wire
Connect black wire
Connect white wire
Connect green wire
Connect red wire
Connect white wire
Connect green wire, jumper 17 and 18
Jumper to 17
Connect red wire
Connect black wire
Connect clear wire
Connect black wire
Connect purple wire
Connect coax center wire
Connect clear wire
Connect white wire
Connect black wire
Connect orange wire
Connect red wire
Connect green wire
Connect blue wire
II-9

Advertisement

Table of Contents
loading

Table of Contents