Thermo Scientific Orion 2117HL User Manual page 91

High level chloride analyzer
Table of Contents

Advertisement

Troubleshooting
Keypad Test
Default
Operator Action
Keypad Test
Press all keys one at a time in any order
No operator action needed
The system will enter the measure mode
if the keypad test is passed
Thermo Scientific Orion 2117HL High Level Chloride Analyzer User Guide
• The main display will show 0 and the marquee will display KEYPAD
TEST PRESS ALL KEYS ONE AT A TIME.
• Press all eight keys one at a time in any order.
* After the first key is pressed Error E033 will be displayed if a new key is
not pressed within 10 seconds.
Scrolling Marquee
KEYPAD TEST
PRESS ALL KEYS ONE AT A TIME
KEYPAD TEST
PRESS ALL KEYS ONE AT A TIME
Main Display
Notes
0
Display will show
After the first
numbers 1 through 8
key is pressed
as keys are pressed
error E033 will
be displayed
if a new key
is not pressed
within 10
seconds
VI-8