Thermo Scientific Orion 2117HL User Manual page 71

High level chloride analyzer
Table of Contents

Advertisement

Calibration
Step
Operator Action
23
Press
to accept the new slope and
E
values
0
or
Press
to abort the calibration. The
new slope and E
0
The last saved calibration prior to the
aborted calibration will be used in the
measure mode.
24
Pull the diverter valve out.
Turn the inlet valve on.
Close the calibration port on
the flow cell cap.
25
Press
.
26
System will proceed to measure mode.
Thermo Scientific Orion 2117HL High Level Chloride Analyzer User Guide
Scrolling Marquee
NEW CALIBRATION SLOPE AND E
PRESS ENTER TO ACCEPT OR PRESS
EXIT TO ABORT WITHOUT SAVING
values will not be saved.
PULL DIVERTER VALVE OUT
TURN INLET VALVE ON
CLOSE CAL PORT
PRESS ENTER
PULL DIVERTER VALVE OUT
TURN INLET VALVE ON
CLOSE CAL PORT
PRESS ENTER
Actual temperature reading
Main Display
Calculated Slope
0
and E
0
FLO On
FLO On
Actual concentration
reading
Notes
If
is pressed,
refer to the
Calibration Abort
Steps section.
System enters
measure mode, but
will remain on hold
until the hold timer
expires. Default hold
time is 30 minutes.
IV-9