Thermo Scientific Orion 2117HL User Manual page 69

High level chloride analyzer
Table of Contents

Advertisement

Calibration
Step
Operator Action
9
Press
.
10
When the liquid level in the flow cell
begins to drop, turn the inlet valve off.
Open the calibration port on the flow
cell cap.
11
Press
.
12
Wait 2 minutes.
13
Wait 0 to 8 minutes.
14
Add standard 1 through the
calibration port.
Pipet 0.5 mL of the standard into the flow
cell using the 0.5 mL pipet.
15
Press
.
16
Wait 2 minutes.
Thermo Scientific Orion 2117HL High Level Chloride Analyzer User Guide
Scrolling Marquee
CHECK CAL PORT CLOSED
PUSH DIVERTER VALVE IN
PRESS ENTER
WHEN LIQUID LEVEL BEGINS TO DROP
TURN INLET VALVE OFF
OPEN CAL PORT
PRESS ENTER
WHEN LIQUID LEVEL BEGINS TO DROP
TURN INLET VALVE OFF
OPEN CAL PORT
PRESS ENTER
STARTING CALIBRATION
MEASURING BACKGROUND
WAIT 2 MINUTES
AWAITING DATA STABILITY
ADD STD1
PRESS ENTER
ADD STD1
PRESS ENTER
MEASURING STD1
CIRCULATING
WAIT 2 MINUTES
Main Display
Notes
FILL
FCEL
FLO OFF
FLO OFF
Actual readings
System circulates
flashing
sample. Minutes will
be counted down on
the marquee.
Actual readings
System continuously
flashing
evaluates the
data until a stable
condition is met.
If there is a calibration
error, the appropriate
error message will be
displayed.
Add Std1
For calibration
standard 1, use the
29,600 ppm high level
chloride standard
(Cat. No. 211740HL).
Add Std1
Actual readings
System circulates
flashing
sample. Minutes will
be counted down on
the marquee.
IV-7