Glossary - Thermo Scientific Orion 2117HL User Manual

High level chloride analyzer
Table of Contents

Advertisement

General Information

Glossary

Thermo Scientific Orion 2117HL High Level Chloride Analyzer User Guide
Refer to Figure I-3.
Inlet Valve – Accepts the sample stream via 1/4 inch NPTF connector. The
operator must supply the sample with a pressure between 14 and 100 psig.
Inlet Filter – 60 micron stainless steel filter traps particulate matter in the
sample stream.
Bypass/Needle Valve – Used to redirect flow in the bypass system.
Pressure Regulator – Adjusts flow on the incoming sample stream.
Flow Meter – Measures the sample flow rate.
Check Valve – Prevents the backflow of sample.
Restrictor Tube Assembly – Used in conjunction with the pressure
regulator to lower downstream pressure.
Flow Cell Manifold – Directs sample flow in and out of the flow cell.
Flow Cell – Contains the chloride sensing electrode, reference electrode and
ATC probe.
Diverter Valve – Allows the flow cell reservoir to fill during calibration by
forming a closed-loop system.
Chloride Electrode – Senses chloride ions in the sample stream and
produce an electrical potential dependent on sample concentration.
Reference Electrode – Provides a constant reference potential and
completes the measurement circuit.
Pressurized Reference Electrode Filling Solution Bottle – Provides
constant flow of electrolyte solution through reference electrode for
maximum stability.
ATC Probe – Measures the sample temperature and inputs data to
microprocessor for automatic temperature compensation (ATC).
Calibration Port – Allows introduction of standards to the sample reservoir
during calibration.
Air Pump – Used to mix the sample during both measurement and
calibration as well as provide air to pressurize the reference electrode filling
solution bottle.
I-8