Yearly .Preventative .Maintenance - Thermo Scientific Orion 2117HL User Manual

High level chloride analyzer
Table of Contents

Advertisement

Analyzer Maintenance
Yearly Preventive
Maintenance
Electrodes
Diverter Valve O-rings
valve stem
O-rings (2)
retaining ring
valve stem
Flow Cell Assembly
Thermo Scientific Orion 2117HL High Level Chloride Analyzer User Guide
1. Replace the Thermo Scientific Orion reference electrode
(Cat. No. 100057). Refer to Chapter II, Installing a New Reference
Electrode for step-by-step instructions.
2. Replace the Thermo Scientific Orion chloride sensing electrode
(Cat. No. 210017). Refer to Chapter II, Installing a New Chloride
Electrode for step-by-step instructions.
3. Replace the calibration solutions (Cat. No. 211740HL)
Dispose of all old solutions according to the MSDS instruction. MSDS
files can be downloaded at www.thermo.com/processwater.
Valve Stem O-rings
1. Remove the two screws that hold the valve stem assembly in place.
2. Gently pull on the valve stem to remove.
3. Slide off the two old O-rings and replace them. It is not necessary to
remove the retaining ring.
4. Place the assembly back into the manifold block and tighten screws.
Note: Do not use O-ring grease.
manifold
face
Manifold Face O-rings
1. Remove the flow cell manifold assembly.
2. Remove the three O-rings from the face of the flow cell manifold.
3. Insert the new O-rings on the face of the flow cell manifold.
4. Replace the flow cell manifold assembly.
Figure V-1
V-4