Analyzer Operation
Default
Operator Action
TEST (cont'd)
To Test SOUt:
4-20
Press
4-20 and rLY settings and select 4-20
Press
Press
SOUt and tOUt settings and select SOUt
Press
Actual
low value
displayed
Press
Actual
high value
displayed
Press
to the main setup mode
Press
specific menu options in the setup mode
Press
and relays (rLY)
Thermo Scientific Orion 2117HL High Level Chloride Analyzer User Guide
/
to scroll between
to test 4-20 outputs
/
to scroll between
to display the SOUt low value
to display the SOUt high value
to accept the test and return
to return to the channel
to test the mA outputs (4-20)
Scrolling Marquee
SELECT mA OR RELAY
SELECT TEMP OR SENSOR
OUTPUT
SELECT TEMP OR SENSOR
OUTPUT
mA SENSOR OUTPUT LOW
VALUE
mA SENSOR OUTPUT HIGH
VALUE
CH1 or CH2
(flashing)
TEST
(flashing)
Main Display
Notes
tESt
4 - 20
(flashing)
4 - 20
tOUt or SOUt
(flashing)
4 - 20
SOUt
(flashing)
LO
4.0 mA are sourced at
output terminal 9
# . #
HI
20.0 mA are sourced
at output terminal 9
#. #
SEL
Display for two
channel analyzer only
SCrn
Action for two
channel analyzer only
SEL
SEL CH1 or SEL CH2
on main display of
SCrn
two channel analyzer
III-30