Analyzer Preparation
Terminal Connector Location
Terminal Connector
Thermo Scientific Orion 2117HL High Level Chloride Analyzer User Guide
Figure II-5
5. Wire the power cable to the green screw terminal connector from the
options kit. Select correct terminal for hot conductor depending on line
voltage. Refer to Figure II-5 for terminal connector location.
6. Plug the terminal connector into the power supply. Refer to
Figure II-3.
7. Select the correct fuse from the fuse kit. Install by inserting the fuse in
the fuse holder and secure it using the twist and lock method. The fuses
are clearly labeled with the appropriate voltages for your system. Refer
to Figure II-3. Refer to the table below for fuse selection.
AC Voltage
115V
230V
Fuse Rating
200mA, 250V, Fast Acting
100mA, 250V Fast Acting