Performing A Dka Calibration - Thermo Scientific Orion 2117HL User Manual

High level chloride analyzer
Table of Contents

Advertisement

Calibration
Performing a
DKA Calibration
Step
Operator Action
1
Make sure that the analyzer is in the
measure mode.
2
Press
to initiate the calibration.
3
If the calibration password is active,
enter the calibration password by pressing
/
to set the first digit,
to move to the next digit,
to set the values of the remaining
digits and
to move through the
remaining digits.
Press
to accept the password.
4
5
Press
/
channel that will be calibrated and press
to begin the calibration.
6
7
Press
/
press
to accept the setting.
8
Check that the calibration port on the flow
cell cap is closed.
Push the diverter valve in.
Thermo Scientific Orion 2117HL High Level Chloride Analyzer User Guide
To begin a DKA calibration, the system starts from the measure mode.
The calibration port must be closed for proper flow cell pressure.
The value of the calibration standards, volume of the calibration
standards and the volume of the flow cell must be entered in the setup
mode prior to performing a calibration. Refer to Chapter III, Use of
the Setup Mode and Chapter IV, Calibration Setup.
Scrolling Marquee
Actual temperature reading
ENTER PASSWORD
/
CH1 or CH2
(flashing)
CH1 or CH2
to select the
(flashing)
DKA or OFFL
(flashing)
DKA
to select DKA and
(flashing)
CHECK CAL PORT CLOSED
PUSH DIVERTER VALVE IN
PRESS ENTER
Main Display
Notes
Actual concentration
reading
CAL appears as mode
PASS
# # # #
CAL
Displayed for two
channel analyzer
CHnL
CAL
Action required for
two channel analyzer
CHnL
CAL
tYPE
CAL
tYPE
FILL
FCEL
IV-6