Analyzer Operation
Using Password
Protection
Thermo Scientific Orion 2117HL High Level Chloride Analyzer User Guide
The default password is 0000 – indicates password protection has not
been activated.
System password: Management secured password protection of setup
mode and calibration process.
Calibration password: Operator secured password for protection of
calibration process only.
If password(s) are activated:
System prompts operator to enter system password:
Marquee: ENTER PASSWORD
Main display top: PASS
Main display bottom: 0000 (flashing)
Correct password – Allows operator to enter setup mode for custom
programming options.
Incorrect password – Password incorrect or not entered correctly.
System password:
Marquee: SYSTEM PASS INCORRECT
Main display: E035
Calibration password:
Marquee: CAL PASS INCORRECT
Main display: E034
Verify password and re-enter it.
If password(s) are de-activated:
System enters the setup mode at the last setup menu option used by
the operator.
Marquee: Flashes current menu option
Main display: SEL SCrn
III-4