Principles Of Operation - Thermo Scientific Orion 2117HL User Manual

High level chloride analyzer
Table of Contents

Advertisement

General Information

Principles of Operation

2117HL Schematic
Thermo Scientific Orion 2117HL High Level Chloride Analyzer User Guide
The sample enters the Thermo Scientific Orion 2117HL chloride analyzer
and passes through the inlet valve, bypass/needle valve, inlet filter, pressure
regulator, flow meter and into the restrictor tubing. The sample then passes
through the flow cell manifold into the flow cell as air is introduced from
the air pump to ensure proper mixing and fast response. The sample passes
the chloride sensing electrode, reference electrode and ATC probe and then
flows into an atmospheric drain via the diverter valve.
Figure I-1
Power
4-20mA Output
Safety Drain
for Overflow
Analyzer
1
2
3
4
5
Calibration
Calibration Level
Measure Level
Cal
Val
Valve ve
Valve
al
Sample
Flow Cell Block
Valve Positions
Calibration Mode - PUSH
Flow Cell Manifold
Sample/Measure Mode - PULL
Drain
Pressurizes Reference
Filling Solution Bottle
Restrictor Tubing
Air Filter
Air Inlet
Pressurizes Flow Cell
to Initiate Siphoning
Port
Regulator
Bypass / Needle
Valve and Filter
Inlet Valve
Main Feed
K E Y
Electric
Fluid Path
Fluid Restrictor Tubing
Air Path
Reference Solution Path
Overflow Path
Check
Valve
Valve
I-3

Advertisement

Table of Contents
loading

Table of Contents