Appendix
Specifications
Thermo Scientific Orion 2117HL High Level Chloride Analyzer User Guide
Chloride Measurement
Range
Accuracy (with DKA calibration)
Resolution
Response Time
Units Displayed
Temperature Compensation
mV Measurement
Range
Resolution
Relative Accuracy
Temperature Measurement
Range
Resolution
Relative Accuracy
Failure Detection
ATC Probe
LED Status Indicator
Green
Yellow
Red
75 ppm to 1000 ppm
± 10%
2, 3 or 4 significant digits
50% of final reading within 2 minutes of injecting a
standard solution
ppb, ppm (auto ranging)
Automatic
± 1999.9 mV
0.1 mV
± (0.5 mV + 0.1%)
-10 to 120 °C
0.1 °C
± 0.5 °C
Manual compensation
30 K thermistor
OK
Warning
Failure
A-3