Calibration
Calibration
Abort Steps
Step
Operator Action
Calibration Abort Exit Steps – If
Exit 1
Exit 2
Press
to abort the calibration.
or
Press
to return to the last screen
used in the calibration sequence prior to
when
was pressed. Continue to
follow the calibration steps according to
the marquee and main display directions.
Exit 3
Exit 4
Press
to abort the calibration.
Exit 5
Pull the diverter valve out.
Turn the inlet valve on.
Close the calibration port on the
flow cell cap.
Exit 6
Press
.
Exit 7
No operator action needed.
The system will enter the measure mode
and remain on hold until the hold timer
expires (default hold time is 30 minutes).
Thermo Scientific Orion 2117HL High Level Chloride Analyzer User Guide
The system allows the operator to abort the calibration or return to the
calibration when
The following table lists the calibration abort steps with their marquee
display and required operator actions.
Scrolling Marquee
is pressed at any point during the calibration, proceed with the following steps
ARE YOU SURE?
PRESS ENTER TO ABORT OR PRESS
LAST SCREEN TO BACKUP
ARE YOU SURE?
PRESS ENTER TO ABORT OR PRESS
LAST SCREEN TO BACKUP
CALIBRATION ABORTED
PRESS ENTER
CALIBRATION ABORTED
PRESS ENTER
PULL DIVERTER VALVE OUT
TURN INLET VALVE ON
CLOSE CAL PORT
PRESS ENTER
PULL DIVERTER VALVE OUT
TURN INLET VALVE ON
CLOSE CAL PORT
PRESS ENTER
Actual temperature reading
is pressed at any point during the calibration.
Main Display
SUrE
?
SUrE
?
CAL
Abrt
CAL
Abrt
FLO On
FLO On
Actual concentration
reading
Notes
IV-10