Specifications - Thermo Scientific 2210AM Manual

Ammonia analyzer
Table of Contents

Advertisement

Specifications

Measurement Performance
0.01 to 2 mg/L ammonia as N
0.05 to10 mg/L
Measuring Range
0.5 to 50 mg/L
50 to 500 mg/L
±5% or 0.01 mg/L (as N) for 0.01 to
50 mg/L whichever is greater
Accuracy
±10% (as N) for 50-500 mg/L for
5°C to 40°C
Measurement
Less than 20 minutes per analysis
cycle time
for ranges up to 500 mg/L
Less than ±3% of reading or
Repeatability
±0.05 mg/L, whichever is greater
from 0-500 mg/L
Limit of
0.01 mg/L
Detection
1 pack of reagents every 45 days
with 2 hour cycle time up to
10 mg/L using the LR reagent kit.
Reagent
For ranges above 10 mg/L1 pack of
Consumption
reagents every 45 days with
2 hour cycle time using the HR
reagent kit.
Method
Salicylate colorimetric method
Sample Requirements
Sample Flow
50 to 1000 mL/min
Sample Pressure
5 psig max
Sample Supply
Continuous
Sample
Temperature
5°C to 40°C (41°F to 104°F)
Range
Sample
1/4 inch OD flexible tubing –
Inlet/Outlet
Polypropylene or similar material
Connections
3/8 inch OD flexible tubing –
Drain Tubing
Polypropylene or similar material
Sample Streams
One
2 |
General Information
Sample Requirements for Analyzer Inlet
Inlet Water
TDS less than 170 mg/L
Quality - TDS
Inlet Water
TSS less than 30 mg/L
Quality - TSS
Inlet Water
pH within the limits of 4 to 10
Quality - pH
Inlet Water
High color sample consult factory
Quality
If sample composition does not meet above
requirements a sampling conditioning panel is
required.
The presence of organic nitrogen compounds may
interfere with the measurement of Ammonia.
Environmental
Ambient
Operating
5°C to 45°C (41°F to 113°F)
Temperature
Maximum
90% at 40°C (104°F)
Humidity"
Construction
NEMA12 (Electronics cabinet),
Enclosure Integrity
NEMA4X (Fluidics cabinet)
Enclosure
27.5" x 16" x 6.4"
Dimensions
(698 mm x 406 mm x 162 mm)
Shipping Weight
18 kg (40 lbs)
Electrical Data and Control
Power
100-240 V AC, 110 W, 50/60 Hz
Requirements
Two 0/4-20 mA – Direct or Reverse
Current Loops
Acting (Isolated). Maximum 900
ohm load.
Orion 2210AM Ammonia Analyzer