Electrical service

specifications

Thermo Fisher Scientific
Specifications
The following table lists the specifications for electrical service. Contact
your local Thermo Fisher Scientific service representative if you have
questions about the requirements. If you are not sure that your power
lines meet these requirements, contact our customer support for
information about power audits. Locate your power source within 3 m
(10 ft) of the instrument rear panel.
Item
Electrical service
Nominal operating AC
input
Nominal operating
current
Maximum startup
AC input & current
Line frequency
Line disturbances
Noise
System components such as the computer, printer and monitor are not
included in these specifications. See the documentation that came with
those components for specifics.
Specification
3-wire
90 V to 264 V
1.7 A at 115 V
1.0 A at 230 V
15 A at 115 V
30 A at 230 V
47 Hz to 63 Hz
Sags, surges or other line disturbances
must not exceed 10 % of input voltage
(even for a half cycle)
< 2 V (common
mode)
Antaris II and Antaris MX Site and Safety Information
live
neutral
earth ground
< 20 V (normal mode)