Product Overview

Intended use

Orion™ 8010cX Ammonia Analyzer is a product developed
by Thermo Fisher Scientific for on-line detection of ammonia
nitrogen (ammonia nitrogen, ammonium radical and
ammonia). Its application fields include the following aspects:
1.On-line detection of conventional wastewater: including
on-line detection of pollution sources, inlet and outlet
detection of sewage treatment facilities, etc.

Specifications

Table 2. Specifications of Orion 8010cX Ammonia Analyzer
Measuring performance
Orion 8010cX Ammonia Analyzer
Range of measurement
Accuracy
Repeatability
Limit of detection (LOD)
Resolution ration
Analysis cycle
Test method
2.On-line detection of surface water: including on-line
detection of water sources, lakes, reservoirs, etc.
3.Industrial process on-line control: for sites where industrial
processes need on-line detection.
4.On-line detection of drinking water: quality control of
disinfection process and on-line detection of drinking water.
For applications not mentioned above, contact your local
dealer or Thermo Fisher Scientific Inc.
Range 1: (0.02 - 2) mg/L;
Range 2: (0.1 - 15) mg/L;
Range 3: (0.5 - 30) mg/L;
Range 4: (2 - 100) mg/L;
Range 5: (30 - 500) mg/L;
Five ranges in total
Range 1: (0.02 - 2) mg/L: 3% ±0.04 mg/L of the reading
Range 2: (0.1 - 15) mg/L: 3% ±0.1 mg/L of the reading
Range 3: (0.5 - 30) mg/L: 3% ±0.2 mg/L of the reading
Range 4: (2 - 100) mg/L: 3% ±0.3 mg/L of the reading
Range 5: (30 - 500) mg/L: 5% ±4 mg/L of the reading
Range 1: (0.02 - 2) mg/L: 3% or ±0.02 mg/L
Range 2: (0.1 - 15) mg/L: 3% or ±0.05 mg/L
Range 3: (0.5 - 30) mg/L: 3% or ±0.1 mg/L
Range 4: (2 - 100) mg/L: 3% or ±0.3 mg/L
Range 5: (30 - 500) mg/L: 3% or ±1 mg/L
Range 1: (0.02 - 2) mg/L: 0.02 mg/L
Range 2: (0.1 - 15) mg/L: 0.1 mg/L
Range 3: (0.5 - 30) mg/L: 0.5 mg/L
Range 4: (2 - 100) mg/L: 1 mg/L
Range 5: (30 - 500) mg/L: 5 mg/L
0.001 mg/L for the reading < 100mg/L
0.01 mg/L for the reading ≥ 100mg/L
Continuous and periodic measurement (start time can be
set)
Salicylic acid colorimetry
Product Overview |